Can I perform additional amplification of libraries prepared with the ThruPLEX DNA-Seq Kit if the concentrations or yields of the libraries are insufficient?

Can I perform additional amplification of libraries prepared with the ThruPLEX DNA-Seq Kit if the concentrations or yields of the libraries are insufficient?

Yes, ThruPLEX DNA-Seq libraries can be further amplified without adding extra reagents after storage (4°C for up to 6 hours, or –20°C for up to 7 days). Please refer to the ThruPLEX DNA-Seq Kit User Manual for the library processing workflow and instructions on performing additional amplification cycles.

DNA-seq products

related articles